: This report details the case of a 41-year-old woman who was diagnosed with insulinoma. As the patient developed severe life-threatening hypoglycemia, we introduced Dexcom G4 Platinum (DG4P), a modern continuous glucose-monitoring system (CGM). The algorithm of the sensor glucose (SG) values of CGM is based on patients with diabThis report details the case of a 41-year-old woman who was diagnosed with insulinoma. As the patient developed severe life-threatening hypoglycemia, we introduced Dexcom G4 Platinum (DG4P), a modern continuous glucose-monitoring system (CGM). The algorithm of the sensor glucose (SG) values of CGM is based on patients with diabetes; therefore, we evaluated the accuracy of DG4P in this patient. The mean absolute relative differences and absolute differences between SG of DG4P and self-monitoring of blood sugar values were 10.8%±8.3% and 6.8±5.7 mg/dL, respectively, in the hypoglycemic region, which verifies DG4P's accuracy. DG4P was found to be useful for monitoring hypoglycemia not only in patients with diabetes but also in those with insulinoma.etes; therefore, we evaluated the accuracy of DG4P in this patient. The mean absolute relative differences and absolute differences between SG of DG4P and self-monitoring of blood sugar values were 10.8%±8.3% and 6.8±5.7 mg/dL, respectively, in the hypoglycemic region, which verifies DG4P's accuracy. DG4P was found to be useful for monitoring hypoglycemia not only in patients with diabetes but also in those with insulinoma.